Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems

Company
Beckman Coulter Inc.
Recall Initiated
January 2, 2025
Posted
February 5, 2025
Recall Number
Z-1112-2025
Quantity
6,994 US; 10,067 OUS
Firm Location
Brea, CA

Reason for Recall

Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time required to achieve a passing calibration curve.

Distribution

Worldwide distribution - US Nationwide and the countries of Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Denmark, France, Germany, Greece, Hungary, India, Ireland, Italy, Kazakhstan, Korea, Republic of, Malaysia, Mexico, Peru, Philippines, Poland, Puerto Rico, Romania, Slovakia, South Africa, Spain, Switzerland, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Viet Nam.

Lot / Code Info

UDI-DI 15099590736002 Lots 339062, 439811

Root Cause

Under Investigation by firm

Action Taken

An Urgent Medical Device Recall letter was sent during the week of December 30, 2024, to the affected customers via first class mail and email. If a passing calibration curve can be generated using the Access PCT reagent lots listed above, no action is needed. If repeated calibration failures with a failure code of Max Iterate are obtained using the Access PCT reagent lots listed above, discontinue using the affected reagent lot number.