Helix Elite" Inactivated Standard: Inactivated SARS-CoV-2 Whole Virus (Pellet), Multi-analyte, unassayed quality control material, Catalog Number HE0065N

This recall is currently active, issued March 14, 2025. It was issued by Microbiologics Inc.

Company
Microbiologics Inc
Recall Initiated
February 20, 2025
Posted
March 14, 2025
Recall Number
Z-1371-2025
Quantity
6 units
Firm Location
Saint Cloud, MN
Official Source
View on FDA website ↗

Reason for Recall

The A549 human cell target (human cells for sample adequacy control) may give a late Ct value.

Distribution

US Nationwide distribution in the states of TX. UK.

Lot / Code Info

UDI-DI: 10845357043563; Lot Number: HE0065-29

Root Cause

Process control

Action Taken

An Urgent Medical Device Recall notification letter dated 2/20/25 was sent to customers. INSTRUCTIONS: 1. REVIEW lab procedures to understand how this information affects your usage 2. DISCARD affected product. 3. COMPLETE the response form provided. 4. RETURN the response form to recall@microbiologics.com 5. KEEP this letter for your records. 6. NOTIFY other facilities or customers to whom the product may have been distributed or transferred. 7. CONTACT Microbiologics if a replacement kit is needed. Microbiologics is dedicated to ensuring the success of your quality control processes. Please contact our Recall Support team at 320.229.7080 or recall@microbiologics.com with questions or concerns. Collect calls may be made.