Z-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision of a physician or other certified healthcare professional.
This recall is currently active, issued November 21, 2024. It was issued by Zyno Medical Llc.
- Company
- Zyno Medical Llc
- Recall Initiated
- October 18, 2024
- Posted
- November 21, 2024
- Recall Number
- Z-0506-2025
- Quantity
- 34,994 units
- Firm Location
- Natick, MA
- Official Source
- View on FDA website ↗
Reason for Recall
Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.
Distribution
This product is distributed only in the United States.
Lot / Code Info
Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF Lot Code: Lots: 201016280, 200515888, 200406416, 230920212, 200413400, 20040660, 200716968, 201016720, 20071632, 201211740, 201208256, 210319640, 210709900, 201211360, 2101221000, 210913560, 210913200, 231008352, 220205400, 220204548, 2211141000, 230206980, 230620800, 230420548, 230420136, 230525452, 230620200.
Root Cause
Device Design
Action Taken
Zyno Medical notified consignees on 10/18/2024 via email. Consignees were instructed that the updated notification supersedes current labeling and includes the following instructions: To maintain the health and runtime of the battery, it is recommended to plug the battery into AC Mains in for at least 6 hours every 30 days to fully charge the battery. Additionally, for scenarios where the Z800 pump is left plugged into AC Mains for extended periods of time, it is recommended to unplug for up to 5 seconds and re-plug the Z800 pump back into AC Mains at least every 30 days to activate the internal battery charger and ensure an expected operating time on battery power. Customers were also asked to provide acknowledgement of receipt.