EMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: 342-14-705
- Recall Initiated
- January 5, 2026
- Posted
- February 2, 2026
- Recall Number
- Z-1233-2026
- Quantity
- 19
- Firm Location
- Austin, TX
Reason for Recall
942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery.
Distribution
US: TX, IA, MI, MS, MA, RI, NJ, PA, NY, MO, OK, PR, AL, CA
Lot / Code Info
UDI-DI: 00888912167529. Lot: 139T1075A. Expiration: 03-12-2026
Root Cause
Labeling Change Control
Action Taken
On 1/5/2026, recall notices were emailed to customers who were asked to do the following: 1) This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 2) Contact the firm to arrange for affected device return. 3. Complete and return the acknowledgement and response form via email to Lesli.Helmick@enovis.com Customers with questions can email the firm at productsafety@enovis.com