RecallDepth

Mojo Non Vented Full Face Mask with Headgear, All Sizes

This recall is currently active, issued April 1, 2024. It was issued by Sleepnet Corporation.

Company
Sleepnet Corporation
Recall Initiated
March 1, 2024
Posted
April 1, 2024
Recall Number
Z-1352-2024
Quantity
11,874 units
Firm Location
Hampton, NH
Official Source
View on FDA website ↗

Reason for Recall

Update to contraindications and warning language due to CPAP masks containing magnets.

Distribution

Worldwide distribution - US Nationwide and the countries of AUSTRIA, BAHRAIN, BRAZIL, COLOMBIA, COSTA RICA, ECUADOR, EL SALVADOR, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, INDIA, IRAQ, IRELAND, ISRAEL, ISREAL, ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, MALAYSIA, MEXICO, MOROCCO, NORWAY, POLAND, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SULTANATE OF OMAN, SWEDEN, SWITZERLAND, THE NETHERLANDS, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VIETNAM, YEMEN.

Lot / Code Info

All UDI-DI; All Lot Numbers.

Root Cause

Device Design

Action Taken

The firm emailed consignees a notification titled URGENT: MEDICAL DEVICE RECALL, dated 3/1/24. Consignees were instructed to stop using recalled masks if the user or anyone in close physical contact with the mask has an active medical implant or metallic implant that will interact with magnets and contact their mask provider for a replacement mask that does not contain magnets. Consignees that do not have an active medical implant or metallic implant that will interact with magnets do not have to cease use. If a consignee is unsure whether or not to cease use, they should consult with their physician or manufacturer of the implant. Consignees are to continue using masks according to the updated instructions and labeling if patients or people in close proximity to them do not have implanted metallic medical devices or metallic objects in the body. Consignees with any questions are to contact Himabindu Koppusetty via email at hkoppusetty@sleepnetcorp.com or by phone at 1-800-742-3646x6611 from Monday through from, 8:00 AM to 4:00 PM.

More recalls by Sleepnet Corporation

View all recalls by this company →