Stryker ACM (Advanced Cement Mixing): 180-Gram Cement Cartridge with Breakaway Femoral Nozzle-Allows the user to mix bone cement with a constant high vacuum through the process Catalog Number: 0206-530-000

This recall is currently active, issued September 19, 2023. It was issued by Stryker Corporation.

Company
Stryker Corporation
Recall Initiated
August 7, 2023
Posted
September 19, 2023
Recall Number
Z-2597-2023
Quantity
224 units US; 718 units OUS
Firm Location
Portage, MI
Official Source
View on FDA website ↗

Reason for Recall

Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site

Distribution

Nationwide Foreign: Austria Belgium Canada France Germany Hong Kong Italy Malaysia Poland Singapore South Korea Spain Taiwan Thailand UK

Lot / Code Info

GTIN: 04546540857880 Lot Numbers: 22332012 22336012 22333012 22355012 23052012

Root Cause

Under Investigation by firm

Action Taken

Stryker issued Urgent Medical Device Recall letter on 8/07/23. Letter states reason for recall, health risk and action to take: Actions needed 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if affected product is no longer in inventory, to the email Instruments.Recalls@stryker.com. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A replacement order will be shipped to customers upon receipt of the recalled product. 5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. If you have any questions about this matter, contact Stryker Instruments by email at Instruments.Recalls@stryker.com.