RecallDepth

MR 7700 Magnetic Resonance (MR) system, Product Numbers 782120 and 782130 (Upgrade to MR 7700)

This recall is currently active, issued August 11, 2023. It was issued by Philips North America.

Company
Philips North America
Recall Initiated
June 12, 2023
Posted
August 11, 2023
Recall Number
Z-2400-2023
Quantity
601 total
Firm Location
Cambridge, MA
Official Source
View on FDA website ↗

Reason for Recall

During installation of the SmokeDetector Interlock, a connecting cable, between the SmokeDetector Interlock and the host PC, may not have been properly installed. This connection is used for service diagnostics and does not impact the functionality of the system and/or the SmokeDetector Interlock system.

Distribution

US Nationwide. Global Distribution.

Lot / Code Info

Product Number: 782120: UDI-DI: 00884838104112; Serial Numbers: 65041 65031 65030 65025 65023 65019 65013 65009 6500865007 65006 65005 65004 65003 65002 65000 65020; Product Number 782130: UDI-DI: 00884838104402; Serial Numbers: 4530145298 45007 45297

Root Cause

Under Investigation by firm

Action Taken

An URGENT Medical Device Correction Notice dated 6/5/23 was sent to customers. Actions that should be taken by the customer /user in order to prevent risks for patients or users: 4.1. Philips has installed a SmokeDetector Interlock feature on all the systems listed in Appendix A. Post this notice near the affected MR system(s) for ease of reference. Circulate this notice to all users of this device so that they are aware of the product issue and associated hazard/harm until this issue has been resolved. Please complete and return the attached updated customer response form to Philips promptly and no later than 30 days from receipt of this letter. Philips is requesting that even if you acknowledged a previous version of this letter that you also respond with an acknowledgment to this updated letter. 4.2. The SmokeDetector Interlock feature is integrated with the MR System and powers off the relevant part of the system(Gradient Amplifier) when smoke is detected in order to prevent further development of smoke and/or fire. If the SmokeDetector Interlock has detected smoke and has powered off the Gradient Amplifier: Remove patient from the system according to the instructions for use because power to the tabletop will still be active Scanning is disabled until Philips service has checked the system. Do not attempt to continue scanning. Immediately inform Philips Service DO NOT attempt to restart or perform a power cycle of the system 4.3. If a user experiences smoke, fire, or severe image quality issues (image artifacts caused by electrical disturbances): Immediately stop scanning and evacuate the patient from the Examination Room. Check Examination Room for a developing fire. Ifa developing fire is sensed, adhere to established fire emergency procedures, which may include switching off power to the complete system and/or removing the magnet field by using the Emergency Magnet Off button Scanning is disabled until Philips service has checked the s

More recalls by Philips North America

View all recalls by this company →