Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000

This recall has been completed (originally issued December 22, 2023).

Recall Initiated
November 15, 2023
Posted
December 22, 2023
Recall Number
Z-0616-2024
Quantity
584 units
Firm Location
Glenwood, IL
Official Source
View on FDA website ↗

Reason for Recall

MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.

Distribution

US and Australia, Brazil, Canada, China, France, Germany, India, Japan, Mexico, Rep of Korea, Taiwan, Thailand, Viet Nam

Lot / Code Info

All serial numbers.

Root Cause

Other

Action Taken

Landauer Issued an URGENT - Medical Device Recall notice to its consignees on 11/15/2023 via email. The notice explained the issue, potential risk, and requested return of the device.