Cold Form Wrap: Back, REF: DP163CT03-BLK-L,XL, and Shoulder, REF: DP163CT02
- Company
- Djo, Llc
- Recall Initiated
- September 25, 2025
- Posted
- November 26, 2025
- Recall Number
- Z-0722-2026
- Quantity
- 1,594
- Firm Location
- Carlsbad, CA
Reason for Recall
The IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures capable of causing a second degree burn.
Distribution
US Nationwide distribution in the states of WI, TN, TX, MD, GA, UT, WA, MI, NC, NE, CA, MA, CO, SC, OH, IN, MN, NM, PA, NY, IL, FL, NJ, VA, AZ, VT, CT, OK.
Lot / Code Info
REF/UDI-DI/Lot: DP163CT02/00888912164382/190523, 171221. DP163CT03-BLK-L,XL/00888912164481/150221, 180821
Root Cause
Labeling Change Control
Action Taken
Per recall notice, customers were asked to do the following: Users should immediately discontinue all further use of this product and discard any opened/used units. If you have unopened, unused units in your possession, please contact firm Customer Service at (844) 331-7767. You will be provided with a return authorization number and shipping label that is required for proper return and identification of the device. All users must complete the acknowledgement and response form and return to productsafety@enovis.com This notice must be distributed to all those within your organization and to any other organization, consignee, or user where these products were transferred. If you have any questions you can reach the firm via email at productsafety@enovis.com