Custom Tubing Kit: REF: K04-00417, K04-00963, K04-01029

Company
Merit Medical Systems, Inc
Recall Initiated
June 5, 2025
Posted
August 12, 2025
Recall Number
Z-2294-2025
Quantity
2779
Firm Location
Jordan, UT

Reason for Recall

High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.

Distribution

US Nationwide distribution in the states of OH, CA, SD, VA, AZ, MT, TX, MI, FL, TN, PA, NJ, MN, ND, IA, LA, UT, KY, IL, OK, ME, IN, NY, NM, SC, AL, ID, CT, WI, GA, MA, NC, AR, WA.

Lot / Code Info

REF/UDI-DI/Lot(Expiration): K04-00417/00884450046630/I3114667(12/10/2027), I3147550(1/28/2028); K04-00963/00884450046760/H3117053(12/14/2027), H3138590(1/20/2028), H3154108(2/10/2028), H3154786(2/10/2028), H3162244(2/19/2028), H3166789(3/2/2028), H3173526(3/23/2028); K04-01029/00884450148938/H3117059(12/17/2027), H3154112(2/10/2028), H3159017(2/19/2028), H3166777(3/9/2028), H3181994(4/1/2028);

Root Cause

Process control

Action Taken

On 5/5/2025, recall notices were emailed to customers who were asked to do the following: 1) Identify affected devices and quarantine, discontinue use and distribution. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the customer response form. 4. Return all affected devices to the firm. 5. Complete, and return the customer response form via email to response@merit.com If you have questions, contact the firm's customer service at RESPONSE@merit.com or via phone at +1 800 356 3748, Hours: 6 am to 6 pm MST, Mon-Fri