3M Ranger Blood/Fluid Warming High Flow Set Catalog Numbers 24355, 24370

Company
3M Company
Recall Initiated
April 10, 2025
Posted
September 18, 2025
Recall Number
Z-2483-2025
Quantity
530,470 units
Firm Location
Saint Paul, MN

Reason for Recall

Field action to clarify the flow rates and inlet fluid temperatures, along with location of the outlet temperature for Ranger Blood/Fluid Warming High Flow Sets.

Distribution

Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Belgium, Bosnia and Herzegovina, Canada, Canary Islands, China, Costa Rica, Croatia, Czechia, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Korea, Luxembourg, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, UAE, UK, and Vietnam.

Lot / Code Info

All lots with a manufacturing date after March 2022.

Root Cause

Device Design

Action Taken

An Urgent Medical Device Correction letter was sent to all their customers on 4/21/2025 to inform them that Solventum [formerly 3M Company] clarified flow rates related to the inlet fluid temperature. The device instructions for use (IFU) will be updated to reflect these changes. Customers are informed that there have been no changes to the design or composition of these products that prompted this notification. Customers are instructed to continue to use the products as indicated, considering the updated information provided in the letter. If the device is used at a flow rate outside of the updated information provided, there is a potential risk for the device to administer the fluids at a lower temperature than indicated and cause a decrease in patient temperature. Under the dire circumstances of a massive rapid infusion where high flow rates outside of the updated ranges indicated below may be used, there is a potential for hypothermia. Customers with additional questions or need to report an adverse event, are instructed to contact the Solventum [formerly 3M Health Care] Customer Helpline at 1-800-228-3957, option 6.

More recalls by 3M Company

View all recalls by this company →