AESCULAP ELAN 4 FIXED DURAGUARD LONG. Model Number: GB943R.

This recall is currently active, issued October 2, 2025. It was issued by Aesculap Ag Am Aesculap-Platz 1 Tuttlingen Germany.

Company
Aesculap Ag Am Aesculap-Platz 1 Tuttlingen Germany
Recall Initiated
August 18, 2025
Posted
October 2, 2025
Recall Number
Z-0027-2026
Quantity
3 units
Official Source
View on FDA website ↗

Reason for Recall

Mislabeling. The ELAN 4 FIXED DURAGUARD STANDARD was labeled as "Long" and the ELAN 4 FIXED DURAGUARD LONG was labeled as "Standard."

Distribution

US distribution to Missouri. OUS International distribution to Norway, Portugal, France, Great Britian, Czech Republic, Germany, Turkey, Brazil, Japan, Chile, China, Hong Kong, Indonesia, India, Iran, Mexico, Malaysia, Philippines, Saudi Arabia, Thailand.

Lot / Code Info

Model Number: GB943R, UDI-DI Number: 04046963805784. Serial Numbers: 4519, 4528, 4533.

Root Cause

Process control

Action Taken

Asculap notified consignees with an "Urgent Field Safety Notice" dated 8/18/2025 via email. Consignees were instructed to identify affected units, arrange for their return, and complete and return the response form provided with the letter. If you have any question, call 484.240.8373 or email:allison.longenhagen@bbraunusa.com.