LIFEPAK 1000 Part / Catalog Numbers / UDI-DI code: 320371500023 / 99425-000025 / 885074698953 320371500229 / 99425-000025 / 885074698953 331706400001 / 99425-000025 / 885074698953 LIFEPAK 1000: The defibrillator is to be used in AED mode only on patients who are in cardiopulmonary arrest. The patient must be unresponsive, not breathing normally, and showing no signs of circulation. ECG monitoring is for use on conscious and unconscious patients of all ages for the purpose of ECG rhythm recognition and heart rate monitoring.

Company
Physio Control, Inc.
Recall Initiated
July 22, 2025
Posted
September 2, 2025
Recall Number
Z-2498-2025
Quantity
3 units
Firm Location
Redmond, WA

Reason for Recall

Due to required inspections not being performed on products/units that have gone through servicing.

Distribution

U.S. Nationwide distribution in the states of CO and MT.

Lot / Code Info

Part / Catalog Numbers / UDI-DI code / Serial Number: 320371500023 / 99425-000025 / 885074698953 / 35942190 320371500229 / 99425-000025 / 885074698953 / 41012118 331706400001 / 99425-000025 / 885074698953 / 48988273

Root Cause

Under Investigation by firm

Action Taken

On 07/22/2025, the firm sent via certified mail an "URGENT: Medical device recall" Letter to customers informing them that, Stryker discovered a subset of work orders performed between August 10th 2022 and October 30th 2023 with Performance Inspection Procedures (PIP) recorded as PASS without the corresponding objective evidence properly documented in the work order to ensure the devices are performing as intended. Customers are instructed to: 1. Immediately check their inventory to locate the affected product as listed in the letter. 2. On the business reply form, customer need to indicate if their devices have been serviced since October 30th 2023 (after that time, if affected device have undergone Preventative Maintenance through Stryker, Stryker has verified that the PIPs were completed and passed in accordance with the required standards. For these devices No additional action is required. 3. Return business reply form. Upon receipt, Stryker will contact customers to arrange for corrective action on the device if service has not been performed. 4. There are no additional actions that users can and should take. Continue to follow the Operating Instructions and perform the daily checks and user tests. Tests can be found in the chapter "Maintaining the Equipment' subsection "General Maintenance and Testing" For Questions or concerns, contact Customer Service at +1 800 787-9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time, Monday thru Friday or email medtechsup@stryker.com