RecallDepth

ELEOS COLLAR STEM, CEMENTED, FLUTED, 15MM X 120MM. Model Number: HC-15120-03M. Component of Limb Salvage System with BioGrip.

Company
Onkos Surgical, Inc.
Recall Initiated
December 22, 2023
Posted
May 23, 2025
Recall Number
Z-1818-2025
Quantity
23 units
Firm Location
Parsippany, NJ

Reason for Recall

Mislabeling of 13mm Modular Segmental Stem package (HC-13120-03M) which included an implant that was 15mm in diameter (HC-15120-03M) instead of being 13mm as stated on the package. The corresponding 15mm cemented stems were mislabeled as size 13mm cemented stems.

Distribution

US Nationwide distribution in the states of California, Kentucky, and Wisconsin.

Lot / Code Info

Model Number: HC-15120-03M. UDI-DI: B278HC1512003M0. Serial Numbers: P230274-1XX (XX represents 23 parts in sequential order from 101-123).

Root Cause

Labeling Change Control

Action Taken

ONKOS Surgical notified their distributors on about 12/22/2023 via email. They were instructed to segregate and quarantine the product immediately, complete and return the "Product and Distribution Information" form within 2 business days and return any existing implants on hand within 2 business days after signing the acknowledgement. They were also instructed that they will receive replacement implants and to place them in their modular segmental collar stem implant packs.

More recalls by Onkos Surgical, Inc.

View all recalls by this company →