Palindrome Precision HSI Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.
This recall is currently active, issued May 30, 2025. It was issued by Mozarc Medical Us Llc.
- Company
- Mozarc Medical Us Llc
- Recall Initiated
- April 23, 2025
- Posted
- May 30, 2025
- Recall Number
- Z-1886-2025
- Quantity
- 95,938 units
- Firm Location
- Mansfield, MA
- Official Source
- View on FDA website ↗
Reason for Recall
Potential breach of sterile barrier packaging.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada & Mexico.
Lot / Code Info
Model No.8888128459P, 8888119369P, 8888133509P, 8888123409P; UDI: 20884521157870, 20884521157733, 20884521157948, 20884521157801; Lot/Serial No. 2133600044, 2201700104, 2203300067, 2228000065, 2228000067, 2228000077, 2234800091, 2234800133, 2234800161, 2301100103, 2302500152, 2302500153, 2302500156, 2302500158, 2305300130, 2318500118, 2318500120, 2318500143, 2330300100, 2330300102, 2332700278, 2332700279, 2335200415, 2335200416, 2402400082, 2402400083, 2402400084, 2405800231, 2418000151, 2424200129, 2426300194, 2427700220.
Root Cause
Packaging
Action Taken
An URGENT: MEDICAL DEVICE RECALL notification dated April 2025 was mailed to consignees on 4/23/25. This notification recommends that clinicians follow facility-specific policies and procedures for routine assessment of the hemodialysis access device for patency, function, and efficacy, as well as for monitoring patient status. Consignees are instructed to immediately quarantine and discontinue use of affected devices in inventory. Consignees are to complete the electronic Customer Confirmation form to facilitate obtaining a Return Goods Authorization number for recalled devices to be returned for credit or replaced. Product purchased through a distributor can be arranged through the distributor. Consignees with any questions about the Customer Confirmation form can contact IQVIA by phone at 1-256-680-9942 or by email at medtronic-fa1486@iqvia.com. Consignees with questions about the recall can contact Medtronic Customer Care at 877-211-1850.