restor3d Cordera Hip System. ACETABULAR LINER GRP C 36MM LIP. Model Number: HDL-060-C22L-000101 Acetabular Liner for Cordera Hip System

This recall is currently active, issued April 15, 2025. It was issued by Conformis Inc..

Company
Conformis Inc.
Recall Initiated
March 25, 2025
Posted
April 15, 2025
Recall Number
Z-1569-2025
Quantity
16
Firm Location
Wilmington, MA
Official Source
View on FDA website ↗

Reason for Recall

The hip cup liner that was packaged with the incorrect labeling. HDL-060-C22S-000101 from Lot 1788258 contained a hooded liner, but it was labeled to have a standard liner.

Distribution

US Nationwide distribution in the states of CA, FL, NV, TN.

Lot / Code Info

Model Number: HDL-060-C22L-000101 UDI-DI: M572HDL060C22L011. Lot: 1788258

Root Cause

Labeling Change Control

Action Taken

On about 03/25/2025, Conformis notified consignees via email with letter titled "URGENT: MEDICAL DEVICE RECALL." Consignees were instructed to examine inventory, quarantine any affected units on hand, and make arrangements to return the product. Additionally, consignees were asked to complete and return the response portion of the letter.