RecallDepth

SURGICEL FIBRILLAR 1INX2IN(2.5CMX5.1CM), Product Number: 1963

This recall is currently active, issued July 17, 2024. It was issued by Ethicon Sarl, A Johnson & Johnson Company Rue Du Puits-Godet 20 Neuchatel Switzerland.

Company
Ethicon Sarl, A Johnson & Johnson Company Rue Du Puits-Godet 20 Neuchatel Switzerland
Recall Initiated
June 11, 2024
Posted
July 17, 2024
Recall Number
Z-2407-2024
Quantity
39,010 units (All US)
Official Source
View on FDA website ↗

Reason for Recall

Some foil pouches were found torn open, impacting the sterile barrier which was traced to a specific machine used during product packaging.

Distribution

Worldwide distribution - US Nationwide and the countries of India and Japan.

Lot / Code Info

Product Number: 1963; UDI/DI: 10705031003132 (primary), 20705031003139 (box of 10); Lot numbers:

Root Cause

Process control

Action Taken

On Jun 11, 2024, URGENT: MEDICAL DEVICE RECALL (REMOVAL) letters were sent to consignees. ACTION REQUIRED: 1. Examine your inventory immediately to determine if you have the lot numbers from the product codes listed in Attachment 1 on hand and quarantine such product(s). If you have product subject to this recall, please maintain a copy of this notice with the quarantined product and keep a copy for your records. 2. Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 3. Complete the Business Reply Form (BRF) (Attachment 3) confirming receipt of this notice and fax or email to Sedgwick at 855-207-8799 or ethicon8137@sedgwick.com within three (3) business days. Please return the BRF even if you do not have product subject to this recall. 4. Customers are required to return unused SURGICEL" FIBRILLAR" Absorbable Hemostat subject to this recall that are in inventory immediately. Credit will be issued to customers who ordered directly from Ethicon. Credit or replacement product as appropriate will be provided to customers who ordered through a distributor. To receive credit or replacement product, customers must return product subject to this recall no later than September 30, 2024. Any non-affected product and any product returned after the date specified will not be eligible for credit or replacement product. 5. To return product subject to this recall, photocopy the completed BRF, place it in the box with the product, and affix the pre-paid authorized shipping label included with the recall notification letter. Ethicon will pay for the shipping charges only if the authorized label is used. Extra shipping labels may be obtained by calling Sedgwick at 855-620-5696. Your account number and ma

More recalls by Ethicon Sarl, A Johnson & Johnson Company Rue Du Puits-Godet 20 Neuchatel Switzerland