The D*Clot HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system consists of an outer sheath through which a rotatable radiopaque spiral shaft extends ending in an eccentric blunt distal tip. Once activated, the shaft rotates at approximately 10.000 RPMs to break thrombus into smaller particles, which can be simultaneously aspirated using a dedicated vacuum syringe. The catheter is a stand-alone system and is powered by a 4.5 VDC alkaline battery pack in the control unit.

This recall is currently active, issued June 26, 2024. It was issued by Mermaid Medical A/S Frydensbergvej 25 Stenlose Denmark.

Company
Mermaid Medical A/S Frydensbergvej 25 Stenlose Denmark
Recall Initiated
May 1, 2024
Posted
June 26, 2024
Recall Number
Z-2213-2024
Quantity
190 devices
Official Source
View on FDA website ↗

Reason for Recall

Internal product testing failed bubble leak testing (ASTM F2096), which indicates the sterility barrier may not be fully intact.

Distribution

Worldwide - US Nationwide distribution in the states of OH, MO, CO, NC, TX, MA, GA, FL, IL, IN, MD, PA, VA, and the country of Chile.

Lot / Code Info

Unique Device Identifier: 15711055004333; Lot codes: S2306001; Model Number: T10106065

Root Cause

Under Investigation by firm

Action Taken

The consignees and distributors were sent out the notification via email on 05/01/2024. The letter instructs the consignees to: 1. Quarantine and Discontinue Use of Affected Product: Identify, segregate, and discontinue use of the affected product in your inventory. 2. Complete and Return the attached Response Form: Complete the attached Response Form and return to Mermaid Medical by May 10, 2024 via email or by post. 3. Distributors/ If You Further Distributed Affected Product: Forward this communication to any customers that were distributed affected product. Instruct any of your customers to complete the response form and return to Mermaid Medical. Once they complete and return the response form, Mermaid Medical will provide instructions on how to return or destroy any affected product remaining in their inventory.