Box as labeled: Regard, Carter-Thomason CloseSure System XL, REF 733776, Quantity 1 Box of 3 Systems. Kit as labeled: Regard, Carter-Thomason CloseSure System, REF 733776, Contents: One (1) Carter-Thomason Suture Passer, One (1) 5 mm Pilot Guide, One (1) 10/12 mm Pilot Guide.

Company
Coopersurgical, Inc.
Recall Initiated
August 29, 2023
Posted
February 5, 2024
Recall Number
Z-0985-2024
Quantity
127 boxes, 3 kits per box (381 kits total)
Firm Location
Trumbull, CT

Reason for Recall

Incorrect component descriptions within the "Contents" section on the Tyvek lid for individual kit trays.

Distribution

US Nationwide distribution in the state of MO.

Lot / Code Info

UDI-DI: 60888937027053, lot# 326728.

Root Cause

Labeling Change Control

Action Taken

An Urgent Medical Device Field Safety Notice dated August 29, 2023 was sent to the one distributor receiving shipments of the affected product lot. Customers are advised to complete the acknowledgement form, Affected product is to be returned to CooperSugical for credit. Contact CooperSurgical at +1-203-601-5200 extension 3300 during normal operating hours of 09:00 - 17:00 M-F EST, email Recall@coopersurgical.com.