Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine. Material #11515088

This recall is currently active, issued July 27, 2023. It was issued by Siemens Medical Solutions Usa, Inc.

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
June 5, 2023
Posted
July 27, 2023
Recall Number
Z-2240-2023
Quantity
9 units
Firm Location
Malvern, PA
Official Source
View on FDA website ↗

Reason for Recall

Incorrect glue was applied on the plexiglass sheet on the single tank cover assembly could potentially fall off and the ingress of liquid may occur, the risk of electric shock exists (e.g., ingress of a large amount of liquid following the plexiglass dropping off, at the same time the user or operator happens to touch the liquid on the cover which is connected to the internal live parts).

Distribution

Worldwide - US Nationwide distribution in the states of FL, GA, KS, MN, NC, NY and the countries of Austria, Bangladesh, Brazil, Canada, China, Croatia, El Salvador, France, Germany, Indonesia, Italy, Japan, Jordan, Lebanon, Luxembourg, Mauritius, Mexico, Monaco, Peru, Philippines, Poland, Romania, Slovakia, South Africa, Spain, Switzerland, Taiwan, Ukraine, United Kingdom.

Lot / Code Info

UDI-DI :04056869148243 US S/N: 30001 30002 30008 30013 30029 30031 30032 30033 30034

Root Cause

Nonconforming Material/Component

Action Taken

Siemens issued Customer Safety Advisory Notice via AX015/23/S on 6/6/23. Letter states reason for recall, health risk and action to take: Siemens will correct the hardware error via Update Instruction AX012/23/S. This measure will replace the current single tank cover of all affected systems. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein