Sensor Nitinol Wire with Hydrophilic Tip .038 3CM FLEX STR/150CM UPN: M006670309 1box 5 UPN: M0066703090 single unit

Company
Boston Scientific
Recall Initiated
March 9, 2020
Posted
April 1, 2020
Recall Number
Z-1731-2020
Quantity
5,565
Firm Location
Marlborough, MA

Reason for Recall

An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device

Distribution

Nationwide Foreign: Europe, Middle East and Africa, the Americas (North America and South America, Asia Pacific and Japan

Lot / Code Info

Lot Numbers: 24501748, 24506585, 24507390, 24507747, 24517205, 24517209, 24517211, 24526341, 24576484, 24576569, 24577900, 24579133, 24580102, 24583255, 24585905, 24586401, 24616721, 24617337, 24618334, 24724936, 24732742, 24737184, 24739802, 24790717, 24805720, 24860825, 24868932

Root Cause

Component change control

Action Taken

Boston Scientific issued Customer Notification on March 9, 2020 stating reason for recall, health risk and action to take: cease distribution or use of any remaining product affected by this removal should cease immediately; Verify by product lot/batch number in the product table whether any product within your inventory is affected. If so, indicate on your Verification Form the quantity of units from each lot/batch that you will bereturning. As some of these products within these lots/batches are sold as 5-packs, it is important that all reported quantities represent the actual number of single units being returned and not the number of cartons/boxes or multi-packs; If you identify any product from the affected lots/batches within your inventory, please segregate the product immediately and return it to BSC in accordance with the enclosed instructions. If you are a distributor, please note that the depth is to the hospital level and this notification should be forwarded to your customers. If you are a facility that has sent products to another hospital within your network, please ensure that this notification is forwarded to them.