Canon Aquilion One, Model TSX-305A/3 V8.3 with FIRST 2.1

Company
Canon Medical System, Usa, Inc.
Recall Initiated
March 25, 2019
Posted
April 10, 2020
Recall Number
Z-1726-2020
Quantity
35 devices
Firm Location
Tustin, CA

Reason for Recall

Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

Distribution

Distribution was made to AK, AR, CA, FL, IA, ID, IL, KS, MA, MD, MI, MN, MO, MT, NC, ND, NH, NY, OH, OK, OR, PA, TN, TX, VT, WA, WI, and WV. There was government and military distribution but no foreign distribution.

Lot / Code Info

Canon Aquilion ONE GENESIS Edition, Whole-Body X-Ray CT Scanner, Model TSX-305A - Serial number 3LA1652001, 3LA17Y2020, 3LB1872025, 3LA1762010, 3LA1772012, 3LA1762009, 3LB1882026, 3LA1752007, 3LA1762008, 3LA1772013, 3LA1732006, 3LA1772015, 3LA1772014, 3LA16Y2003, 3LA1792017, 3LA1782016, 3LA1712005, 3LB1832021, 3LA16Z2004, 3LA17Y2019, 3LB1842022, 3LA1772011, 3LB1852024, 3LA17X2018, 3OB1852023, 6LB18Y2013, 6LB1892008, 6LB1862002, 6LB1882007, 6LB1852001, 6LB1892009, 6LB18Y2011, 6LB1872005, 6LB1872006, and 6OB1862003.

Root Cause

Software design

Action Taken

Recall letters dated 3/25/2019 were issued on 3/25/2019 via USPS return receipt mail.