AmniSure ROM Test, Material Nos. FMRT-1-25-US (US Distribution, 25 tests/box), FMRT-1-10-US (US Distribution, 10 tests/box), FMRT-1-25-CA (Canadian Distribution; 25 tests/box)

Company
Qiagen Sciences Llc
Recall Initiated
February 21, 2020
Terminated
October 5, 2020
Recall Number
Z-1636-2020
Quantity
1654 kit boxes
Firm Location
Germantown, MD

Reason for Recall

An image on the associated polybag contains an incorrect rinse time of 30 (thirty seconds). The correct rinse time is 1 (one) minute.

Distribution

Nationwide domestic distribution. Foreign distribution to Canada.

Lot / Code Info

FMRT-1-25-US: Lots 563010976, 563011620, 563011164, 5660397, 5660552  FMRT-1-10-US: Lot 563011618  FMRT-1-25-CA: Lot 563011195

Root Cause

Employee error

Action Taken

On February 21, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed of the incorrect rinse time on the product's labeling, and reminded of the correct rinse time of 1 (one) minute. Customers were asked to immediately share the contents of the customer letter with all users of the AmniSure ROM Test in their facility to ensure awareness. Test users should refer to the instructions for use (IFU) included in each kit box. The firm is not asking for product to be returned. Customers were asked to complete and return an attached response form.