Teleflex MEDICAL RuSCH One piece Male External Medium Catheter W/O Tape, REF A1500M

Company
Teleflex Medical
Recall Initiated
February 27, 2019
Terminated
September 6, 2023
Recall Number
Z-0797-2020
Quantity
13104 devices
Firm Location
Morrisville, NC

Reason for Recall

The device label incorrectly states that the device is sterile. This product is an externally applied male condom-type catheter designed for the management of male incontinence as an alternative to incontinence pads. In normal use, the device is neither expected nor required to be sterile.

Distribution

ME,NC, WA, MO, NJ, FL, CAL, TX, OH, IL, MA, GA, FL, GA, Canada

Lot / Code Info

GTIN 24026704549843 Lot Numbers: 74A1800577, 74G1802121, 74A1801939, 74H1700894, 74B1800363, 74J1700864, 74B1801604, 74J1702489, 74B1802746, 74C1801178, 74K1702257, 74L1700074, 74C1801294 , 74L1701701, 74C1802132, 74M1700121, 74G1800047, 74M1701150, 74G1800891, 74M1701769, 74G1801166

Root Cause

Under Investigation by firm

Action Taken

The firm initiated the recall by letter on 02/27/2019. The notice stated that the device label incorrectly identified the products as sterile. Although incorrectly labeled as sterile, there is no requirement for sterility for its use. The firm requested that the distributors notify their customers. The firm did not request the return of the units, only to attach the notice to any recalled product in stock.