RecallDepth

BA-800M Chemistry Analyzer Part Numbers BA81E-PA00002 and BA81E-PA00003, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring))

Company
Mindray Ds Usa, Inc. D.B.A. Mindray North America
Recall Initiated
December 9, 2019
Terminated
September 17, 2020
Recall Number
Z-0920-2020
Quantity
7
Firm Location
Mahwah, NJ

Reason for Recall

The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers.

Distribution

The products were distributed to the following US states: MI and TX.

Lot / Code Info

Serial Numbers: W1-6C000006,  W1-71000007,  W1-73000009,  W1-73000010,  XP-68000687T,  W1-71000008,  W1-85000011

Root Cause

Nonconforming Material/Component

Action Taken

The consignee notification was sent certified mail with return receipt verification. The letter identified the affected product, problem and actions to be taken by the Customer/User. Customers were instructed to: Continue to use the system normally and exercise caution when lifting/closing the front and rear cover. Attachment A includes a list of system(s) identified by model and serial number purchased by your company. We recommend that you notify the clients to which you have distributed these systems and advise them to continue normal use with the recommended caution. The gas springs on the systems listed in Attachment A will require replacement and instructions are provided in Attachment B. The parts needed to perform the replacement will be shipped to you within the next several weeks. Type of Action by the Company Mindray will provide replacement gas springs at no cost. Should you require technical assistance while preforming the replacement, please contact Mindrays Technical Support team at 800-288-2121, option 2 (Technical Support), then choose option 3 (IVD Support team).