Arrow Berman Angiographic Catheter 50cc Item Number: AI-07130

This recall has been terminated (originally issued November 15, 2019).

Company
Arrow International Inc
Recall Initiated
November 15, 2019
Terminated
June 19, 2020
Recall Number
Z-0715-2020
Quantity
71 units
Firm Location
Reading, PA
Official Source
View on FDA website ↗

Reason for Recall

Potential leak in the balloon extension line. If a leak was present and went undetected during pre-use testing as outlined in the IFU, this would cause the balloon to deflate slowly following placement. If the leak went undetected by attending clinicians during use, it could result in contrast injection into the vascular or cardiac tissue which could result in serious injury or death

Distribution

U.S. Nationwide, Canada

Lot / Code Info

Lot Number: 16F19H0089

Root Cause

Process control

Action Taken

Arrow International issued letters on 11/15/19 stating reason for recall, health risk, and action to take: immediately discontinue use and quarantine any products with the product codes and lot numbers listed above. 2. To acknowledge receipt of this Urgent Medical Device Recall Notice or to return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1- 855-419-8507, Attn: Customer Service or ema il recalls@teleflex.com. your local sales representative or Customer Service at 1-866-396-2111. .

More recalls by Arrow International Inc

View all recalls by this company →