(1) Argon Medical Devices Stopcocks, Clear 4-Way Stopcock w/Male Luer Lock, REF 041210002A, 25/box, Sterile, Rx; (2) Argon Medical Devices Stopcocks, 1-Way w/Male Luer Lock, REF 041216000A, 25/box, Sterile, Rx; (3) Argon Medical Devices Stopcocks, Double 4-Way Stopcock w/Male Luer Lock, REF 041220001A, 25/box, Sterile, Rx; (4) Argon Medical Devices Stopcocks, Clear 4-Way Stopcock w/Rotating Male Luer Lock, REF 041240002A, Sterile, Rx; (5) Argon Medical Devices Stopcocks, North Carolina Baptist Hospital, Winston-Salem, NC, REF 042597300A, 25/box, Sterile, Rx; and (6) Argon Medical Devices 3 Way Stopcock w/Rotating Adapter (1050psi), REF 196590, 25/box, Sterile, Rx. The firm name on the labels is Argon Medical Devices, Inc., Athens, TX.

This recall has been terminated (originally issued October 16, 2018).

Company
Argon Medical Devices, Inc
Recall Initiated
October 16, 2018
Terminated
June 15, 2021
Recall Number
Z-0290-2020
Quantity
Total number in recall for all products in dist: 49,130 units
Firm Location
Athens, TX
Official Source
View on FDA website ↗

Reason for Recall

The seal formed on the edge of the pouch may be incomplete, which could result in a loss of sterility to the product.

Distribution

Distribution was nationwide. There was military/government distribution. Foreign distribution was made to Canada, Australia, the Netherlands, and Spain.

Lot / Code Info

(1) Argon Medical Devices Stopcocks, Clear 4-Way Stopcock w/Male Luer Lock, REF 041210002A-Lot numbers 11226656 and 11228887;  (2) Argon Medical Devices Stopcocks, 1-Way w/Male Luer Lock, REF 041216000A - Lot numbers 11226187, 11229619, and 11227967;  (3) Argon Medical Devices Stopcocks, Double 4-Way Stopcock w/Male Luer Lock, REF 041220001A - Lot numbers 11228888, 11230598;  (4) Argon Medical Devices Stopcocks, Clear 4-Way Stopcock w/Rotating Male Luer Lock, REF 041240002A - Lot number 11228245;  (5) Argon Medical Devices Stopcocks, North Carolina Baptist Hospital, Winston-Salem, NC, REF 042597300A - Lot number 11232206; and  (6) Argon Medical Devices 3 Way Stopcock w/Rotating Adapter (1050psi), REF 196590 - Lot numbers 11229800 and 11232485.

Root Cause

Equipment maintenance

Action Taken

The recalling firm began issuing letters dated 10/16/2018 on 10/16/2018 via registered mail identifying the issue and the action to be taken by the customer.