VOLUMAT MC AGILIA US, Infusion Pump, REF Z021135, software versions 1.7 and 1.9a Product Usage: The Volumat MC Agilia is a piece of transportable medical equipment intended to be used in healthcare facilities environment by healthcare trained professionals according to hospital protocols on adults, pediatrics and neonatal human patient to administer via a single channel or mounted on a multiple channels rack accessory: Intermittent or continuous delivery of parenteral fluids (solutions, colloids, parenteral nutrition) and medications (including but limited to diluted drugs, chemotherapy) through clinically accepted IV routes of administration). Transfusion of blood and blood derivatives products. Note that for transfusion procedures, use of a special administration set is required. Volumat MC Agilia can be used with the Dose Error Reduction Software (DERS) Vigilant Drug Lib to define the default and upper hard limits for the Bolus Volume, or dose, by drug. The health care professional can then operate the pump for that drug within the upper hard limits defined in the Drug Library.

Company
Fenwal Inc
Recall Initiated
June 24, 2019
Posted
August 12, 2019
Terminated
March 6, 2023
Recall Number
Z-2064-2019
Quantity
9461 units
Firm Location
Lake Zurich, IL

Reason for Recall

The firm is correcting four software anomalies and KVO (keep vein open) end of infusion alarm priority in Software versions 1.7 and 1.9a of Volumat MC Agilia Software and versions 1.0 and 1.1 of Vigilant DrugLib

Distribution

US Nationwide Distribution

Lot / Code Info

All serial numbers distributed in the US

Root Cause

Other

Action Taken

The firm initiated the recall by letter on 06/24/2019. The letter describes the issues and provided supplemental instructions to be implemented until the new software is installed. The firm issued a press release on 06/28/2019.

More recalls by Fenwal Inc

View all recalls by this company →