RecallDepth

Ascension Modular Radial Head (MRH) Radial Head Implant, Catalog Number, MRH-350-04-WW

Company
Ascension Orthopedics, Inc
Recall Initiated
November 16, 2018
Terminated
November 22, 2019
Recall Number
Z-0794-2019
Firm Location
Austin, TX

Reason for Recall

The product was packaged with the Instructions for Use for a different product (Ascension First Choice DRUJ System Partial Ulnar Head Implant).

Distribution

Distributed to accounts in MO. Foreign distribution to Korea and Thailand.

Lot / Code Info

170691

Root Cause

Vendor change control

Action Taken

On 11/16/18, the recalling firm sent out an Urgent Voluntary Medical Device Correction letter via email to affected customers to notify them of the product issue. A second, updated letter was sent on 12/13/18. The firm requests that all recipients please examine their inventory to determine if they have any MRH Radial Head Implants, catalog number, MRH-350-04-WW, from lot number, 170691. Customers: - If you do have any MRH Radial Head Implants, catalog number, MRH-350-04-WW, from lot number, 170691 please discard the First Choice DRUJ System Partial Ulnar Head Implant instructions packaged with the product and use the attached correct IFU provided to you along with this communication letter. Distributors: - If you do have any unused MRH Radial Head Implants, catalog number, MRH-350-04-WW, from lot number, 170691 please return them to Integra. - We also kindly ask you to provide us the final customer list who may have the affected products. Integra will provide them the correct IFU. Complete the attached Acknowledgement Form and return the completed form by email or fax indicating receipt and review of this notification. Keep a copy of the form for your records. Should you have any questions regarding these instructions, please contact Customer Service Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: custsvcnj@integralife.com Outside USA: 1- 609-936-5400; email: INTERNATIONALCS@integralife.com

More recalls by Ascension Orthopedics, Inc

View all recalls by this company →