U2 Total Knee System Tibial Insert, Posterior Stabilized, #3, 11mm Thick Catalogue number 2303-3032 The U2 Total Knee system is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee cannot be satisfactorily balanced and stabilized at the time of surgery.
This recall has been terminated (originally issued November 24, 2017).
- Recall Initiated
- November 24, 2017
- Terminated
- October 9, 2018
- Recall Number
- Z-1128-2018
- Quantity
- 69 pieces
- Firm Location
- Science Park Hsinchu Taiwan
- Official Source
- View on FDA website ↗
Reason for Recall
UOC received a customer compliant report from our distributor on October 31, 2017 which stated that contents of the package do not match the product labeling.
Distribution
Worldwide Distribution - US Distribution and to the countries of: UK, Iran, Greece, Pakistan, Switzerland, France, China, Taiwan, Italy, Korea and South Africa.
Lot / Code Info
Lot number 17B621AX
Root Cause
Labeling mix-ups
Action Taken
Urgent Field safety notice was distributed to customers on November 24, 2017 via e-mail. The letter identified the affected product, problem and actions to be taken. For further questions, please call (949) 328-3366.
More recalls by United Orthopedic Corporation No.
USTAR II Hip System Press-Fit Curved Stem, RHS Ti Plasma Spray for the following reference numbers:...
Feb 29, 2024
USTAR II Knee System : Cemented Curved Stem, RHS Ti Plasma Spray, Reference Numbers:
REF: PE2515310...
Feb 29, 2024
USTAR II Knee System Cemented curved stem, RHS, Non Coated for the following reference numbers:
REF...
Feb 29, 2024