Apollo DRF; Model: 9784220831 Product Usage: General, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures
- Recall Initiated
- December 27, 2016
- Terminated
- March 9, 2021
- Recall Number
- Z-1429-2017
- Quantity
- 65 devices in total
Reason for Recall
Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, PA, Puerto Rico, TN, TX, UT, and W.I
Lot / Code Info
08090531 09090591 10040680 15051450 11020798
Root Cause
Nonconforming Material/Component
Action Taken
Customers were notified of the recall via DHL or similar mail service. The letter identified the affected product, problem and actions to be taken.