OEC 9800, fluoroscopic x-ray system

Company
Ge Oec Medical Systems, Inc
Recall Initiated
July 22, 2016
Terminated
December 19, 2016
Recall Number
Z-2629-2016
Quantity
47 total
Firm Location
Salt Lake City, UT

Reason for Recall

A firmware issue in the supplied LCD workstation monitor(s). Under some circumstances, the monitor(s) may appear dark with no video displayed.

Distribution

Distribution US nationwide and Korea.

Lot / Code Info

monitor part numbers: 5448359-01, 5459219-01, or 5475662-03:89-3853, 82-1593, 8S-1935, 89-3356, 89-1323, 8S-2885, 82-0700, 8S-1708, 82-0420, 82-1998, 8S-2027, 82-7101-MH, 8S-2670, P6-0375-L, 8S-1614, 89-0846, 82-0576, 82-3256, 8S-2202, 89-0711,89-1927, 89-1453, 82-1504, 82-0794, 82-3494, 8S-3125, 8S-1845-RC, 89-3725, 8S-1592, 89-3129,82-1182, 82-0986, 8S-0199, 8S-3613, 82-3748, 82-2335, 8S-2889, 82-0596, 89-3020, 82-1515, 8S-0645, 8S-3614, 82-3925, 89-0416, 89-1363, 82-3258, 8S-0347

Root Cause

Software design (manufacturing process)

Action Taken

On July 22, 2016 GE Healthcare Surgery began distributing an Urgent Medical Device Correction letter to customers via mail to inform them of the issue and provide instructions for immediate mitigation of the issue. The letter was sent to 3 titles at each facility affected: 1.) Facility Administrator, 2.) Director/Manager of Radiology and Radiologists, 3.) Radiology Department. Consignees instructured that local GE Healthcare Field Engineer will contact them to coordinate replacement of the affected monitor(s), and will ensure that system meets specifications.