Zyno Medical Administration Set, Sterile. RX For Use with Zyno Medical Infusion Pumps or Gravity Feed Product Code: B2-70071-DF Product Code: B2-70071-DF-120 (expanded recall) Product Usage: Zyno Medical Administration Set is a device used to administer fluids from a container to a patients vascular system through a needle or a catheter inserted into a vein.
This recall has been terminated (originally issued September 27, 2016).
- Company
- Zyno Medical Llc
- Recall Initiated
- July 29, 2016
- Posted
- September 27, 2016
- Terminated
- July 11, 2017
- Recall Number
- Z-2895-2016
- Quantity
- 78063 sets
- Firm Location
- Natick, MA
- Official Source
- View on FDA website ↗
Reason for Recall
Administration Set potential filter leaking
Distribution
US Nationwide Distribution
Lot / Code Info
B2-70071-DF Lot Codes: 14015279 14026620 14046841 14055191 14055786 14056225 14085743 14085742 15035297 15045941 15045940 15046600 15075275 15075835 15076761 15085107 15086621 15095200 15095647 15097179 15105595 B2-70071-DF-120 lot codes: 15066912 15075253 15075990 15076261 15076304
Root Cause
Process control
Action Taken
Zyno issued Recall letter dated July 26th, 2016 to affected customers. Customers who received affected products were contacted between 7/29/2016 and 8/1/2016. Customers were asked to return unused product for replacement. For questions contact Customer Service Representative, at 508-907-7796, or laryloring@zynomed.com. Expanded Recall: Two recall letters issued on 11/3/2016 to address the newly identified lots. One letter for customers who received the initial recall letter and the other letter for newly affected customers who did not receive the initial recall letter.