Floor Mounting Column TS 7500 U, Material Number 1730732; Floor Mounting Column ARTIS, Material Number 1501878 Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recover from the anesthesia. Also, patient transport on the operating table top from a patient transfer system to the operating theater of from the operating theater to a patient transfer system.

Company
Trumpf Medical Systems, Inc.
Recall Initiated
July 10, 2016
Posted
October 5, 2016
Terminated
March 2, 2017
Recall Number
Z-0042-2017
Quantity
84 units
Firm Location
Charleston, SC

Reason for Recall

Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters.

Distribution

Worldwide Distribution - US including CA, CT, GA, IL, NC, OR, SC, VA, WI, WY, and Internationally to Australia, Azerbaijan, Brazil, Finland, France, Germany, Hungary, Italy, Luxembourg, Norway, Poland, Russia, Slovakia, and Switzerland

Lot / Code Info

Serial Numbers - 101900011, 101900018, 101904491, 101904998

Root Cause

Nonconforming Material/Component

Action Taken

Trumpf Medical sent an Urgent Field Safety Notice dated July 8, 2015, to all affected consignees. The letter indicated that the tables should not be used until an authorized service engineer has exchanged the spindles of the affected operating tables. The letter included an acknowledgement form which is to be returned. The letter also requested a sub-recall if the product had been further distributed. For questions regarding this recall call 843-329-0543.