CDI H/S Cuvette The intended use for the CDI H IS Cuvette is as follows: The CDI H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.
This recall has been terminated (originally issued December 11, 2015).
- Recall Initiated
- December 11, 2015
- Terminated
- October 26, 2016
- Recall Number
- Z-0001-2017
- Quantity
- 140,000 units
- Firm Location
- Elkton, MD
- Official Source
- View on FDA website ↗
Reason for Recall
Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.
Distribution
Worldwide Distribution - USA (nationwide) Distribution.
Lot / Code Info
6912 , 6913 6914, 6932, 6933 6934, 6922, 6923 6924, CV-6912, CV-6913, CV-6914 Cardiovascular Procedure Kits- Various
Root Cause
Other
Action Taken
Terumo (Planned Action) plans to execute the following steps as part of the correction activity: Notify all customers of the potential that a cuvette may disengage from the circuit (probe) leading to a "disconnect" message on the monitor. This notification will be performed via a Customer Notification Letter. For further questions, please call (410) 392-7277.