RecallDepth

CDI H/S Cuvette The intended use for the CDI H IS Cuvette is as follows: The CDI H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.

This recall has been terminated (originally issued December 11, 2015).

Company
Terumo Cardiovascular Systems Corporation
Recall Initiated
December 11, 2015
Terminated
October 26, 2016
Recall Number
Z-0001-2017
Quantity
140,000 units
Firm Location
Elkton, MD
Official Source
View on FDA website ↗

Reason for Recall

Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.

Distribution

Worldwide Distribution - USA (nationwide) Distribution.

Lot / Code Info

6912 , 6913 6914, 6932, 6933 6934, 6922, 6923 6924, CV-6912, CV-6913, CV-6914 Cardiovascular Procedure Kits- Various

Root Cause

Other

Action Taken

Terumo (Planned Action) plans to execute the following steps as part of the correction activity: Notify all customers of the potential that a cuvette may disengage from the circuit (probe) leading to a "disconnect" message on the monitor. This notification will be performed via a Customer Notification Letter. For further questions, please call (410) 392-7277.

More recalls by Terumo Cardiovascular Systems Corporation

View all recalls by this company →