wDR 2.2 Mobile Digital Diagnostic X-Ray System

Company
Sedecal Sa Pelaya, 9 Pol. Pol. Ind. Rio De Janeiro Algete Spain
Recall Initiated
May 23, 2023
Posted
June 30, 2023
Recall Number
Z-2045-2023
Quantity
1061 units

Reason for Recall

There is a software login in issue that may prevent the user from logging in.

Distribution

Distribution throughout United States including Puerto Rico and US Virgin Islands OUS distribution to Albania, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Czech Republic, Dominican Republic, Egypt, Finland, France, Germany, Hungary, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia,Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Martinique, Mexico, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Suriname, Sweden, Switzerland, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam, and Zambia

Lot / Code Info

Model MobileDiagnost wDR: 1) Model Number 712007, Software versions 2.2.0, 2.2.1, 2.2.2, and 2.2.3 2) Model Number 712006, Software versions 2.2.0, 2.2.1, 2.2.2, and 2.2.3 3) Model Number 712004, Software version 2.1 with Windows 10 Upgrade 4) Model Number 712002, Software version 2.1 with Windows 10 Upgrade

Root Cause

Software change control

Action Taken

A recall notification letter was mailed to customers in the United States by the US distributor, Philips, beginning May 23, 2023. Customers are advised to take the following actions: 1) To prevent the system from becoming inoperable, do not reboot or log out of your system until Philips has implemented the software correction. 2) Circulate this notice to all users of the device so that they are aware of the issue. 3) Complete and return the response form to Philips promptly and no later than 30 days from receipt. A Philips representative will contact consignees to schedule a time for a Field Service Engineer to visit the site and implement a software solution to resolve the issue.