Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Model Number 0998-00-0800-55, Part Number 0998-00-0800-XX
- Company
- Datascope Corp.
- Recall Initiated
- May 4, 2023
- Posted
- June 14, 2023
- Recall Number
- Z-1907-2023
- Quantity
- 4 units
- Firm Location
- Mahwah, NJ
Reason for Recall
The digital IFUs provided with the Cardiosave IABP during the software update to version D.00 incorrectly annotated compliance to standard 60601-1-12:2014 within the updated IFU.
Distribution
International distribution in the country of Germany.
Lot / Code Info
UDI-DI: 10607567108414; Serial Numbers: CB286971I8 CB283804G8 CB283867G8 CB283522G8
Root Cause
Under Investigation by firm
Action Taken
An URGENT FIELD SAFETY NOTICE MEDICAL DEVICE CORRECTION notification letter was sent to customer on 5/4/23. Actions to be taken by the customer: " Please locate the CARDIOSAVE Hybrid IABPs with the serial numbers listed in this notice. " Please ensure that all Cardiosave Hybrid Intra-Aortic Balloon Pump users at your facility are aware of this notice. " Please forward this information to all current and potential Cardiosave Hybrid IABPs users within your hospital / facility. " Please contact your local Getinge Service Representative to arrange to have the affected IFU USB drives replaced. " Please complete and sign the attached URGENT FIELD SAFETY NOTICE RESPONSE FORM (page 3) to acknowledge that you have received this notification. Return the completed form to Datascope Corp./Getinge by e-mailing a scanned copy to fieldaction@getinge.com Actions to be taken by Getinge: Getinge will provide your facility with an updated IFU USB drive that contains reference to the vibration testing standard, 60601-1-12:2014. The original USB drive that was provided during the D.00 software update can be destroyed locally. This voluntary correction notification only affects the products listed on page 1; no other products are affected by this voluntary correction. We apologize for any inconvenience this Medical Device Correction may cause. If you have any questions, please contact your Datascope Corp./Getinge representative.