EMPOWR Porous Knee Patella Over Drill, REF: 802-05-104; EMPOWR Knee Patella Drill, REF: 802-05-108

Recall Initiated
February 28, 2023
Posted
April 12, 2023
Recall Number
Z-1384-2023
Quantity
29
Firm Location
Austin, TX

Reason for Recall

Patella drills are marked with incorrect diameter size of 5.6mm, when they should be 5.4mm. Porous patella overdrills are marked with incorrect diameter size of 5.4 mm, when they should be 5.6 mm. If the incorrect drill is selected during surgery, there is risk of the implant loosening, implant not properly fitting, and/or fractured bone.

Distribution

US: TX, IN, MD, CA, KS, AL, HI, WA

Lot / Code Info

REF/UDI-DI/Lot: 802-05-104/00190446238843/388620L01; 802-05-108/00190446365907/388620L03

Root Cause

Process change control

Action Taken

On 2/28/23, field safety notices were emailed to surgical sales agents who were asked to document the quantity of affected devices complete and return the acknowledgement form. Customer service contact, customerservice@enovis.com, was provided to report failures or to request replacements. Customers with questions can phone (914) 450-3931, or email livia.kuruvila@enovis.com.