STA Compact Max fully automatic clinical analyzer, Reference Numbers 58602 and 58612
- Company
- Diagnostica Stago, Inc.
- Recall Initiated
- March 22, 2023
- Posted
- May 10, 2023
- Recall Number
- Z-1577-2023
- Quantity
- 5 analyzers
- Firm Location
- Parsippany, NJ
Reason for Recall
An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error reported by the analyzer (error 13).
Distribution
US Nationwide distribution in the states of IL, MA, MN, NE, TX, and WI.
Lot / Code Info
UDI-DI: 036074505898; Serial Numbers: 5061471 5061474 0000007405 CF70015885 CF70015886
Root Cause
Device Design
Action Taken
Customers were contacted via telephone and an URGENT: MEDICAL DEVICE CORRECTION notification letter beginning 3/22/23. Actions: Stago downgraded your firmware to the previous firmware version as a precaution. Your downgrade has already occurred. Consequently, the risk of this bug only existed during the time period shown above (i.e. from new firmware installation date to the downgrade date). As patient results are interpreted in association with other biological tests, we leave the decision to review previous patient results based on the complete clinical context to your discretion. Our Customer Service team is available to help you carry out your impact analysis. For additional information, please contact your Stago Hotline at 1-800-725-0607.