Newton Wire Guide (LT Taper/10cm Distal Flexibility), Guidewire Reference Part Number/GPN: TSFNA-35-145 G00701 TSCFNA-35-145-3 G00561
- Company
- Cook Incorporated
- Recall Initiated
- August 16, 2022
- Posted
- September 28, 2022
- Recall Number
- Z-1836-2022
- Quantity
- 42 US
- Firm Location
- Bloomington, IN
Reason for Recall
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Lot / Code Info
TSFNA-35-145 G00701 UDI-DI: 00827002007017 Lots/Expiration Date: 14798674 16-06-2027 TSCFNA-35-145-3 G00561 UDI-DI: 00827002005617 Lots/Expiration Date: 14835600 11-07-2027
Root Cause
Under Investigation by firm
Action Taken
On August 16, 2022, Cook Medical notified affected customers of the recall. Customers were instructed to immediately cease all further distribution and use of the affected products. Affected product should be quarantined and returned to Cook Medical to receive product credit. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235