RecallDepth

RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200

This recall has been terminated (originally issued November 1, 2021).

Company
Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland
Recall Initiated
November 1, 2021
Terminated
March 9, 2023
Recall Number
Z-0328-2022
Quantity
15 kits
Official Source
View on FDA website ↗

Reason for Recall

When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range

Distribution

US Nationwide distribution in the states of GA, OR, PA ,WA, West Virginia.

Lot / Code Info

Lot Numbers/Exp Date: 559344 28th June 2022; 572730 28th July 2023. GTIN: 05055273201130

Root Cause

Under Investigation by firm

Action Taken

Randox USA, Distributor contacted the customer via Urgent Medical Device Correction letter email on 11/1/21. The letter states reason for recall, health risk and action to take: Action to be taken: " Discontinue use of and discard any of the above batches immediately. " Review your reagent inventory of these products and assess your laboratories needs for reimbursement for discarded inventory. " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation. If you have any questions or concerns please contact Randox Technical Services.

More recalls by Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland

View all recalls by this company →