Cardiosave Rescue IABP - Product Usage: used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation, Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
- Company
- Datascope Corp.
- Recall Initiated
- March 22, 2021
- Posted
- April 28, 2021
- Recall Number
- Z-1507-2021
- Quantity
- 10087 total
- Firm Location
- Mahwah, NJ
Reason for Recall
There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS).
Distribution
Worldwide distribution - US Nationwide distribution.
Lot / Code Info
All Systems. UDI: 10607567112312 10607567108407 10607567113449
Root Cause
Software design
Action Taken
URGENT MEDICAL DEVICE CORRECTION notification letters dated 3/22/21 were distributed to customers. PLEASE FORWARD THIS INFORMATION TO ALL CURRENT AND POTENTIAL CARDIOSAVE HYBRID and CARDIOSAVE RESCUE IABP USERS WITHIN YOUR HOSPITAL / FACILITY. IF YOU ARE A DISTRIBUTOR WHO HAS SHIPPED ANY AFFECTED PRODUCTS TO CUSTOMERS, PLEASE FORWARD THIS DOCUMENT TO THEIR ATTENTION FOR APPROPRIATE ACTION. Corrective Action: Datascope/Getinge is currently developing a software correction to address this issue. A Datascope/Getinge service representative will contact you to schedule the installation of the updated software. This work will be done at no cost to your facility. Please complete and sign the attached MEDICAL DEVICE CORRECTION RESPONSE FORM (page 5) to acknowledge that you have received this notification. Return the completed form to Datascope/Getinge by e-mailing a scanned copy to cardiosaveripple20.act@getinge.com or by faxing the form to 1-800-506-1644.