Codman Surgical Patties, Part number 801400; SURG PATXRAY 1/2X1/2-200; Lot 6396457

Company
Integra Lifesciences Corp.
Recall Initiated
December 2, 2022
Posted
January 6, 2023
Recall Number
Z-0909-2023
Quantity
168 US; 3 OUS
Firm Location
Princeton, NJ

Reason for Recall

Integra LifeSciences identified that one lot was released with out-of-specification endotoxin results. The worst-case risk for the use of affected Surgical Patties is potentially a minor transient fever.

Distribution

Worldwide distribution - US Nationwide distribution in the states of CA, CO, CT, FL, IN, KY, MA, MI, MO, NC, NJ, NY, OH, PA, TX, VA, WV and the countries of Chile, Australia, New Zealand, France, Romania, United Kingdom, and Ireland.

Lot / Code Info

UDI-DI 10381780514961 Lot Number: 6396457 Expiration date 06/30/2027

Root Cause

Process control

Action Taken

On December 2, 2022, the firm notified customers via "URGENT VOLUNTARY MEDICAL DEVICE RECALL" letters. Customers were instructed to immediately remove units of affected product from service. Customers should complete the Acknowledgement Form and return to the recalling company. Once received, Customer Service will contact the customer to aid them in returning the affected product. Should you have any questions regarding these instructions, please contact Customer Service: Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: custsvcnj@integralife.com