Terumo Capiox NX19 Oxygenator (West Orientation P/N:CXOT281 Exported only to Japan for inclusion into convenience kits that are sold only within the Japanese market

This recall has been terminated (originally issued June 14, 2019).

Company
Terumo Cardiovascular Systems Corporation
Recall Initiated
June 14, 2019
Terminated
June 11, 2020
Recall Number
Z-2104-2019
Quantity
123 units
Firm Location
Elkton, MD
Official Source
View on FDA website ↗

Reason for Recall

Capiox NX19 Oxygenator may contain a non-functioning arterial thermistor component

Distribution

MI Foreign: Japan, Belgium

Lot / Code Info

Lot Numbers: XA21, XA29, XC04, XC18, XD05, XD27

Root Cause

Process control

Action Taken

Terumo issued Urgent Medical Device Notification letter dated June 13 2019 via FedEx mailing executed on 6/14/2019 advising of the problem, health risk and action to take: Terumo CVS advises users to remove all affected product from inventory and return unused product to Terumo CVS following the instructions provided in this notice. Questions or concerns: Terumo CVS Customer Service: 1.800.521.2818 Monday Friday, 8 a.m. 6 p.m. ET Terumo Recall Fax: 1.410.392.7183 Terumo Recall Email: TCVSCustomerResponseElkton@terumomedical.com