RecallDepth

Mindray BeneVision Distributed Monitoring System (DMS), Part No. 115-034413-00

Company
Mindray Ds Usa, Inc. D.B.A. Mindray North America
Recall Initiated
August 30, 2018
Terminated
December 20, 2018
Recall Number
Z-0170-2019
Quantity
5
Firm Location
Mahwah, NJ

Reason for Recall

The firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line at the centralized viewing station and will manifest if more than 64 TD60 transmitters have been connected to a given central station server. Each central station supports only 32 transmitters at a time. For the issue to manifest, the BeneVision DMS needs to be configured to use flexible device assignment, have more than 64 TD60 transmitters, and the workflow must allow 64 TD60s to move around between multiple central stations in the same care area.

Distribution

Distributed to accounts in CA, CO, FL, MA, NY, PA, PR, TX, and WV.

Lot / Code Info

This potential issue affects BeneVision DMS with software version 03.01.00 to 03.03.00

Root Cause

Software design

Action Taken

The firm notified affected consignees on August 30, 2018, via Urgent Medical Device Recall letter. The letter advised customers of the software issue when more than 64 telemetry transmitters have been connected to a central station. Customers were advised to use the system normally, but informed that a Mindray Service Representative would be in contact to arrange for a system software update to be performed at the customer site. Mindrays Technical Support team may be reached at (877) 913-9663 (Option 1), Monday through Friday, 8:30am - 5:30pm ET should you have technical questions. Customers were also asked to complete and return an attached Acknowledgement and Receipt Form.

More recalls by Mindray Ds Usa, Inc. D.B.A. Mindray North America

View all recalls by this company →