daVinci Harmonic ACE Curved Shears

This recall is currently active, issued January 16, 2019. It was issued by Intuitive Surgical, Inc..

Company
Intuitive Surgical, Inc.
Recall Initiated
January 16, 2019
Recall Number
Z-0841-2019
Quantity
43755 inserts
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

During an internal inspection, Intuitive found that the sterile packaging of the da Vinci S/Si Harmonic ACE Inserts may become damaged during transit. The potential damage includes pinholes and /or small openings (less than a few millimeters) on the pouch leading to a potential breach in sterility.

Distribution

US and Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Kuwait, Lebanon, Mexico, Monaco, Netherlands, Pakistan, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Korea, Spain, Taiwan, Turkey, United Kingdom

Lot / Code Info

All Lots

Root Cause

Under Investigation by firm

Action Taken

A Stop Use Notification was sent to affected customers on December 19, 2018. This notification requested customers to return the device. The formal customer recall letter was sent on January 16, 2019.