AVE 2 Birthing Bed
- Recall Initiated
- January 7, 2019
- Terminated
- May 31, 2023
- Recall Number
- Z-0991-2019
- Quantity
- 253 devices
Reason for Recall
LINET has recently become aware of a potential safety issue for the AVE 2 Birthing Bed caused by the power cable breaking at the bed connection or the orange insulation of the power cable may fail causing a possible spark or possible exposure to a live conductor.
Distribution
AL, OK, NJ, OR, WA, CA, MN, LA, AZ, IL, MO, CO, OH, KY, FL, KS, UT, TN, LA, DE, CT, TX
Lot / Code Info
Serial Numbers 4PPB0091 through 4PPB626
Root Cause
Device Design
Action Taken
The firm disseminated their urgent correction notices on 01/07/2019 by mail. The notice discussed the potential safety issue and requested the consignee cease use of the bed if there was damage to the power cable. The firm will be providing a kit that includes a power cable hook and updated instruction for use.