CIVCO Needle Guide, REF 676-144, QTY 24, STERILE EO

This recall has been terminated (originally issued July 30, 2018).

Company
Civco Medical Instruments Co. Inc.
Recall Initiated
July 30, 2018
Terminated
September 16, 2020
Recall Number
Z-2946-2018
Quantity
107 units
Firm Location
Kalona, IA
Official Source
View on FDA website ↗

Reason for Recall

The assembly of the needle guides were found to be either missing adhesive or did not have adequate adhesive to hold the assembled cannula in place on the needle guide.

Distribution

United States, Austria, Belgium, Denmark, France, Germany, Poland, Spain, Switzerland and United Kingdom

Lot / Code Info

UDI 00841436107327, Lot numbers: A018067, A018067, A021026, A021026, A021226, A021226, A021226, A022447, A022447, A023876, A023876,  A025137, A025137, A025137, A025137, A02560, A025606, A025606, A025606, A026609, A026609, A026609, A026609, A026610, A026610, A026610, A028159, A028159, A028645, A029293, A029293, A029293,  A029383, A032036, A032036, A032036, A032036, A032036, A032036, A033139, A033139, A033139, A033139, A033153, A033153, A035009, A035009, A035009, A035010, A035010, A035010, A035010, A035010,  A035568, A035568, A037150, A037231, A037231, A037231, A037585, A037585, A039417, A039417, A040376, A040376, A040839, A041444, A041444, A041444, A041723, A041723

Root Cause

Process control

Action Taken

The firm initiated communication on 07/30/2018 by email and telephone calls. The firm requested return of the recalled product. Questions may be made to the Account Representative by calling CIVCO Medical Solutions, Multi-Modality Imaging at 800-445-6741 or 319-248-6757 7:00 AM - 5:30 PM CST or by email at order@civco.com