Install CD, Spine tools, Plus and S7, Version 25. Model Number 9731958.

This recall has been terminated (originally issued June 26, 2017).

Company
Medtronic Navigation, Inc.
Recall Initiated
May 10, 2017
Posted
June 26, 2017
Terminated
January 8, 2021
Recall Number
Z-2746-2017
Quantity
108
Firm Location
Louisville, CO
Official Source
View on FDA website ↗

Reason for Recall

Software issue related to the StealthStation S7 system and the Synergy Spine application Version 2.1 configured with Spine Tool Install CD version 25. Issue may result in user being unable to navigate the Navigated Elevate Inserter with the StealthAiR Spine Frame during spine surgical procedures.

Distribution

Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Brazil, Germany, Switzerland and Great Britain.

Lot / Code Info

Kits that include affected 9731958 INSTALL CD: INST SET 9734833 UNIVERSAL NAVLOCK, INST SET 9734834 CAPSTONE TRIALS, INST SET 9734835 CLYDESDALE TRIALS, INST SET 9734836 TLIF/DLIF INSERTER, INST SET 9735283 SOLERA 5.5/6.0 DRIVERS, INST SET 9735284 SOLERA NONCAN TAPS, INST SET 9735285 SOLERA ILIAC TAPS, TAPS KIT 9735465 SOLERA AWL-TIP TAPS, REF SET 9735500 SPINE REFRENCING, UPGRD KIT 9735522 SPINE REVISION, INSTR SET 9735607 VERTEX SELECT

Root Cause

Software in the Use Environment

Action Taken

Medtronic sent an Urgent - Product Advisory Notice dated May 10, 2017, to all affected customers. Instructions indicated that a Medtronic representative will be in contact to update the system with version 26 and will remove any copies of the Spine Tools Install CD, version 25. Please quarantine any copies of version 25 until they can be removed. If you have further questions regarding this communication, please contact your local Medtronic Navigation Technical Services Representative, your local Medtronic Spine Sales Representative, or call (800 595-9709.