WECK Visistat 35W, Disposable Skin Stapler, Rx Only, Sterile, Teleflex Medical, NC 27709. The intended use of the Visistat stapler is to close lacerations on the outer layer of the dermis for a wide range of general, thoracic, obstetric, gynecological, orthopedic, cardiovascular, urological, and plastic surgical procedures.

This recall has been terminated (originally issued November 18, 2016).

Company
Teleflex Medical
Recall Initiated
November 18, 2016
Terminated
November 24, 2017
Recall Number
Z-0764-2017
Quantity
96,402 units (9,120 ea US, 87,282 ea OUS)
Firm Location
Research Triangle Park, NC
Official Source
View on FDA website ↗

Reason for Recall

Sterility Deficiency: The packages are not sealed properly and thus the sterility is not assured.

Distribution

Worldwide Distribution- U.S. (Nationwide) and countries of: Australia, Belgium, Canada, China, Germany, Hong Kong, India, Japan, Korea, Malaysia, New Zealand, Singapore and Thailand.

Lot / Code Info

Product Code: 528235, Lot numbers: 73C1600693, 73G1500681, 73H1500255, 73H1500256, 73K1500618, 73L1400006 and 73M1500130.

Root Cause

Packaging process control

Action Taken

The firm, Telexflex Medical, sent an "URGENT MEDICAL DEVICE RECALL NOTIFICATION" letter dated 11/18/2016 to its Consignees/Customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use and quarantine any affected products; complete and return the Recall Acknowledgement Form by fax to 1-855-419-8507, Attn: Customer Service or email to: recalls@teleflex.com (even if you do not have any affected product) and return the product with the completed Recall Acknowledgement Form. A customer service representative will contact you with a RGA # and provide instructions for the return of the product. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-866-246-6990.